We are pleased to present PULSETM 18 with several new applications like PULSE Access, PULSE Array Acoustics Wide Band Holography and PULSE Array Acoustics Moving Source Beamforming, as well as improvements to other application software. The main hardware additions include a 200kHz LAN-XI module and a Wireless Frame for LAN-XI.
This Readme file includes information on Installation, describes all New Applications and New Features, and provides Hints and Tips, Troubleshooting guides, and a list of Resolved Problems.
Recommended PC
We strongly recommend that you update your PULSE installation to the latest major release to ensure that the latest security updates from Microsoft® are supported by your installation.
Warnings:
Windows Vista® and Windows® XP (x64) are not supported.
Microsoft® SQL Server® 2000 is not supported in a workstation environment. It is still possible to use SQL Server 2000 in a network environment.
PULSE Telephone Testing applications using Telephone Testing Software BZ-5137 only work under Windows® XP.
PULSE Wind Turbine Sound Power Determination according to IEC61400-11 Type 7914 does not work under Windows® XP.
Product Support Changes
Please note:
LAN-XI: Backwards Compatibility Broken
With PULSE 18, LAN-XI hardware is updated from PTP (Precision Time Protocol, also known as IEEE1588) v1 to v2. This allows us to utilize new, modern PTP-switches - including one supplied by Brüel & Kjær - but it also means that backwards compatibility is broken. Normally, the newest LAN-XI firmware can run with older versions of PULSE Labshop and Data Recorder Type 7708 Data Recorder, but with PULSE 18, the firmware requires PULSE 18 software.
We are sorry for the inconvenience. The backwards compatibility of LAN-XI is reliant on the standard's capability to support this functionality. It is not possible for us to create a fix.
Installing PULSE
If you are installing from a DVD:
If you have downloaded PULSE:
The software ships with an essential LAN-XI firmware update. Please use the Front-end Setup to ensure firmware update of your LAN-XI modules (after installation, click Start >All Programs > PULSE > Front-end Setup).
The dongle should be disconnected during PULSE installation. After installing PULSE, insert the dongle. The dongle will blink for up to 30 seconds while it is updated. Once updated the dongle should light continuously. See the troubleshooting section if the dongle doesn't update correctly or stop blinking.
Architectural Acoustics Suite
From PULSE 18, the Architectural Acoustics Suite has a new, separate database called "Architectural Acoustics Initial Database". This database contains all the demo data and report templates that greatly enhance post-processing tasks in the Architectural Acoustics Suite.
The change in database structure means that you have to install the new database. When you start the application after software installation, a prompt will show you what you need to do to install the new database. You can keep the prompt open as you perform the steps shown, or make notes and click Close to exit the prompt. If you do not want to see the prompt again, select: Do not show this window again before closing.
You can also choose to skip database installation and start processing data (not recommended). If you use any other database, none of the building acoustic-specific features mentioned above will be included.
To install the new database:
Installing Licenses
To retrieve your license:
If the PC that will be running PULSE is not connected to the Internet, retrieve its Host ID and use another computer that is connected to the Internet to download the license, then transfer the ".lic" files to the computer that will be running PULSE.
Your Host ID can also be found on your License Authorisation Card.
Changing Location of Databases and Filefarm During Installation
During installation of PULSE with custom settings, there is an option to change the location of the databases and the filefarms. When making changes to these settings, please keep the following in mind:
When relocating PULSE File Farm to another folder, the newly specified folder must be named "BKFileFarms" in order to make it work
Backup and remove existing databases before changing locations. Never use mixed locations for databases and filefarms. If databases are already present on the PC, make a backup of every database and then disconnect or delete the databases using PDMTools. Backups can be restored after the custom installation is completed and will be automatically updated with the new database and filefarm location upon restoring.
Installation Troubleshooting
Shortcut to Installation Log File
To view the latest PULSE installation log file, copy
%ALLUSERSPROFILE%\Application Data\Bruel and Kjaer\PULSE latest installation log.lnk
to the address line of Windows® Explorer and press Enter. The link works for all versions of operating system and is updated during install, uninstall, repair, etc.
Antivirus software
If an unexpected dialog appears during installation, disconnect from the Internet and disable your antivirus software's real-time scanner. You should disconnect from the Internet before disabling part of your antivirus software to protect yourself from web-based attacks while the real-time scanning software is disabled. Your antivirus software's options can usually be displayed by clicking on its icon on the task bar.
Error "No socket connection to license server manager"
FLEXlm has been updated to support Windows Server 2008 R2 (x64). Customers using floating licenses should upgrade BKSV.exe, lmgrd.exe and lmtools.exe on their license server to avoid the error from the client software. The files can be found in "Program Files\Common Files\Bruel and Kjaer\FlexLM".
Installation Troubleshooting Page
There are a number of known issues with the installation of PULSE. These have been documented and their respective solutions made available on an Installation Troubleshooting page.
PULSE LabShop
New in Data Recorder Type 7701 (part of Type 7708)
PULSE LabShop now supports Multiple Recorders activated in the system with following specs:
Recording in PTI file format
Recording bandwidth with individual decimation
Bandwidth individually selectable for each recorder
Recording only the selected signal groups plus all auxiliary channels
PULSE LabShop continues to support of DAT/Rec Data Recorder format.
New: PULSE Access Type 7781-N6
PULSE Access provides an entry-level solution to PULSE. It comprises a series of basic analyzers that are the core of any measurement system. These basic analyzers contain all you need for simple sound and vibration tests, operating as stand-alone applications. They can also be expanded to fit larger, industry-specific solutions. By upgrading to PULSE FFT and CPB Analyzers Type 7700, 7770 or 7771, you can quickly gain access to the complete range of PULSE LabShop applications and LAN-XI hardware solutions. This expandability and the continuing development of new PULSE applications and hardware, ensure the safety of your investment now and in the future.
PULSE Access is built in the PULSE LabShop environment:
Delivered with dedicated templates, one for each LAN-XI module
Comes with specifications that meet the market requirements for basic analyzers
PULSE Reflex Core
New in Type 8700
PDM Importer: Direct import to Reflex project browser from a PDM database
Meta-data input/edit component: Dedicated GUI component for listing or modifying meta-data
Four ID lines (meta-data): New meta-data aligned with ID lines of UNV, AFU and ATI file types
New in Types 8702-L and 8702
HATS equalizer element: Apply sound field correction filters for Head & Torso Simulators
AUX signal processing: Axis tagging and signal trigger using auxiliary data
Four ID lines (meta-data): New meta-data aligned with ID lines of UNV, AFU and ATI file types
New in Type 8702 only (not Type 8702-L)
Fully automated processing: Automated import and processing of time files
Input area signal filter: Filter by data name or physical quantity
Result matrix element: New process chain display element using Result Matrix methodology
Function Data Processing: New sub-task enabling function data input to a process chain
Feature Removed from Types 8702-L and 8702
The original Display element has been removed, replaced by the newer Individual Display element, which in general has superior functionality. Saved process chains containing the old display element still work in PULSE 18, but it is highly recommended to update process chains with the newer display element to take advantage of its greater capabilities
New in Type 8706
CPB hold modes: Min./max. hold on individual, single, overall, or overall weighted bands
Changes to Handling of Saved Process Chains
Process chains are now stored in external files and no longer in the database. As a result all saved process chains are available for all new projects in any Reflex application using the Process Chain Editor.
The disk location is \ProgramData\Bruel and Kjaer\PULSE Reflex\ProcessChain\.
Loading old projects including saved chains still works but the chain name is marked as "ready for conversion". The description field of the old chain in the list will display "Please move this chain". If you right-click such a chain you will have the option to "Move chain to external storage". This option will prompt you with a save chain dialog, remove the chain from the current storage and finally create an external file instead. The Import/Export chain option has been adapted to work with external files only. Also, the process chain editor now remembers the name of the last saved chain and reuses this name when saving the chain.
PULSE NVH Simulator
Enhancements
Contribution Analyser: Use one FRF for all operating conditions
Road Creator: Supports GPS files from new Sonoscout
Filtering: Profiled RPM-tracked Notch filter
State Tuner: Option to switch performance models when switching states
Visuals: A selection of up-to-date traffic car visual model
Exterior Sound Simulator: Enhanced sound rendering, including ground reflection with user-defined properties
Time History Displays: In calibrated units or %
New Features
Jury Evaluation: Rating Evaluation now has an optional Detectability button which has many uses, including time-to-detect studies using Exterior Sound Simulator
Visuals: Multi-lane road with five different surfaces specifically developed for road-noise investigations
Mixer: Custom-triggered events (e.g., horn sounds, pitchable loops, sounds triggered by the driving conditions or visual scenario) can be added to a mixer and can be switched on/off in the same way as other sound objects
Recording: If MVR-Triggered recording is enabled, recording is automatically started and stopped to match the mvr file which is currently replaying, e.g., every time you hit a car assembly button in a fixed driving Mixer Test or every time you use the MVR radio button in any free-driving test
SimSound Type 8601-U
SimSound is a new sound generation system for driving simulators. It is used to provide a realistic acoustic scenario to enhance the user’s perception of the virtual environment:
It delivers exceptionally high quality audio signals from complete engine sounds to simple looped sounds
It can be connected to any kind of driving simulator via network or CANbus using a simple plug-in interface
It synthesizes the sounds for the driver’s vehicle, the surrounding traffic and the ambient soundscape in real-time
It is a multi-channel system that can be easily configured to use setups consisting of 2 – 8 speakers plus sub-woofer
It is delivered with a ready-to-go set of vehicle sounds for interior and traffic noise
PULSE Reflex Telephone Testing Type 8770
Type 8770 is a new application for audio performance testing of telephones. The following standards are supported: 3GPP2 CS.0056-0, ITU-T Rec. P.862 and ITU-T Rec. P.863.
Improvements to Array Acoustics Suite
There are new options in PULSE 18:
PULSE Array Acoustics Road Vehicles Moving Source Beamforming BZ-5943
PULSE Array Acoustics Rail Vehicles Moving Source Beamforming BZ-5939
PULSE Array Acoustics Wind Turbines Moving Source Beamforming BZ-5941
PULSE Array Acoustics Wideband Holography BZ-5644
The speed at which the calculations are performed has been increased and the maximum length of time of the recorded files has been increased by a factor of about 5.
Improvements to PULSE Reflex Nastran Interface Type 8718-A
Import of NEi Nastran file.
Improvements to PULSE Reflex Geometry Type 8719
Geometry Creation based on CAD models is now supported in the Geometry Editor task. The Geometry Creation task using our previous geometry tool has been removed. Supported CAD models include:
Curves: Lines, Circles, Circular Arcs, Ellipses, Elliptical Arcs, Hyperbolic and Parabolic curves
Surfaces: Circle, Circular Arc, Ellipse, Elliptical Arc, Hyperbolic and Parabolic surfaces
Solids: Boxes, Cylinders, Hemispheres and Spheres
Extrusion of CAD based geometries. A curve (2D) can be extruded to a Surface (3D) by selecting the curve and dragging it in the height direction. In a similar way, a plane Surface (2D) can be extruded to a Solid (3D).
Element status. During Element creation, a symbol of the Element (Triangle or Quad) illustrates how far the creation process is by coloring selected nodes according to the color definition set in the group Highlight in the Geometry Properties.
Element and traceline colors. The color can now be set immediately after selection of Elements (Triangles or Quads) or Tracelines has been done. Can then be changed in the Elements and Tracelines tables in the Geometry Tree for individual or grouped items.
A number of new Geometry properties especially within the Rendering group.
Additional Select & Edit improvements to the Geometry Tree. This includes:
Multiple selection and editing of table entries using SHIFT and CRTL keys or by dragging cells (Excel-like functionality)
Select All Rows, Deselect All Rows and Invert Selection buttons
Selection of visibility (on/off) of individual coordinate systems, elements and tracelines
Ability to show coordinates in Global or Local coordinate systems
Renumber Nodes functionality including:
Improved link between 3D View and Geometry Tree. Selections made in the geometry 3D View are highlighted in the Geometry Tree tables and vice versa.
Improved usability when decimating FE models to test models. The previously selected Nodes are now given a specific color (default light green) and the currently selected Nodes for the Element or Traceline are given a specific color (default red). After deactivating Element or Tracelines creation, the used Nodes are now highlighted (default light green).
Improved generation of AVI and graphics files. AVI (Save as Video) and graphics files (Save as File) now include all views as displayed on screen whether they are Single, Dual or Quad Views. What you see on screen is what you get in your saved video or graphics file. Also Copy to Clipboard supports this.
Improvements to PULSE Reflex Modal Analysis Type 8720/21
Significant geometry improvements. See Type 8719.
A new ODS Analysis application for frequency-based Spectral ODS post-processing and modal results viewing. Requires only Type 8700 and 8719 licenses. Replaces the previous ability to open up Reflex Modal Analysis Type 8720 without a Type 8720 license for Spectral ODS and modal results viewing.
Improvements to PULSE Reflex Correlation Analysis Type 8722
Significant geometry improvements. See Type 8719.
Improvements to PULSE Sound Power Type 7799
ISO 37454 is now supported. The requirement for a calibration to be made before and after each measurement sequence is not part of the task process. However, calibrations can be made as frequently as required.
Improvements to PULSE Wind Turbine Sound Power Determination according to IEC 61400-11 Type 7914
Type 7914 now supports Edition 3 of IEC 61400-11, which was published in December 2012. However, not all countries are content with this version, so both edition 2.1 and edition 3.0 are included in the new version of Type 7914.
Improvements to PULSE Vehicle Pass-by Test System Type 7788
PULSE LabShop - No data from CAN logger visible
Depending on access rights (no full administrator), CAN BOX installation fails to install in the right locations.
Workaround:
Windows® 7, 32-bit:
Windows® 7, 64-bit:
When installing a new PULSE version, new firmware might have been installed for the CAN BOX. This firmware has to be loaded into the CAN BOX. Open the CANbox configuration by double-clicking on the CANbox icon in the control panel. If new firmware is available, this will be shown as a tickmark on the General tab.
PULSE Data Manager - Crash in Word 2013
When reporting in Word 2013, manipulating PULSE Data Manager tables may cause Word to crash.
PULSE LabShop
In rare cases, PULSE fails to close properly and cannot be restarted. In such cases, go to the Processes tab of the Windows® Task Manager, locate Pulse.exe, right-click and select End Process. Alternatively, restart the PC.
Objects in the PULSE user interface are not sized correctly
Windows® allows you to increase the size of user interface elements (text and other items) to make it easier to read items on the screen. This is not supported by PULSE. For Windows® 7 or 8, open the Display control panel and select Smaller - 100%. For Windows® XP keep the default display settings for text size and display dpi.
Flash Videos
The DVD contains videos, in Adobe® Flash® format, to demonstrate PULSE features. To view these videos the PULSE installer will also install Adobe Flash Player.
The videos offer hints, tips, and step-by-step instructions on many topics. You can also access them on-line from www.bksv.com\UsingPULSE.
The PULSE installation also includes an interactive flash video summary of the LAN-XI product line. This video offers a menu-driven guide to the LAN-XI front-end product line, featuring frame, modules, homepage, accessories, etc. You find the LAN-XI System Information by clicking Getting Started in the DVD menu.
Transducer Database
The transducer database in PULSE 17 is not the same as in PULSE 16. This can lead to limitations in mixed version environments.
Version 16 works with a PULSE 17 database without problems.
Version 17 also works with a PULSE 16 database, but you should be aware that:
Calibrated offset values are silently ignored
If you are using Use Transducer Data in the Channel dialog the only way of specifying a working offset is in the Configuration > Transducer Database > Edit/Add Transducer Type dialog
PULSE LabShop Recorder misses first part of recording when started with signal trigger on 30+ channel systems
Symptom: When using a signal trigger to start a recording with the PULSE LabShop Recorder, some data is missing at the beginning of the recording. This has only been seen on systems with more than 30 channels, but could happen on lower channel systems too.
Solution: The channel used for the signal trigger should be the lowest possible channel according to the Harwdare Setup table or Configuration Organiser.
Zoom in Array Acoustics Post-processing
Manually zooming (Alt + Ctrl + Shift + Left Mouse + Move down) on result maps may cause data points, mesh lines and sound power areas to disappear from the maps.
Workaround: Zoom by right-clicking in the map, selecting Decoration and using the Zoom slider
Communication Error with Front-end: "Samples lost in input module"
In the event of communication error with the front-end, ensure that the netcard in the PC is set to "100 Mbps Full Duplex" instead of "Auto Negotiate" (Link Speed under Advanced Properties).
If this does not solve the problem, try rolling back the driver to an earlier version.
NOTAR: SD Card Formatting
Before making any recordings with a new SD card, it is strongly recommended that you format the SD card using the SD Formatter application provided by the SD Association.
Cards formatted in other devices, including Windows PCs, may not perform well enough to be used with NOTAR, and data loss may occur as a result. NOTAR will indicate this by aborting the recording and displaying Slow SD card and/or Error 402 under Status on the Recorder control page.
Spherical Beamforming calculations calculated with PULSE 15.1 and earlier are always displayed as spherical meshes, even if Unfold Spherical Meshes is selected.
To fix this problem, recalculate the Spherical Beamforming calculation.
Computer with Radeon graphics card freezes
Open the Catalyst Control Center by right clicking on your desktop. Then select crossfire and uncheck Enable CrossFire.
Settings are changed when old files are loaded into Reflex
Some settings may be lost when loading files created in PULSE Reflex 14.1
Changing the transducer in a project might lead to a wrong unit in your data
When changing the transducer on an input in the configuration organiser, the unit might change to V. To avoid this problem you can either change the transducer in the Hardware Table or change the transducer in the Configuration Organiser by removing the current transducer and adding the new transducer.
Noise Source Identification Type 7752: limited use of Edit/copy function
In Noise Source Identification Type 7752 the Edit/Copy function does not work. Use Print Screen instead. The Edit/Copy function works in Array Acoustics Post Processing.
HASP dongle flashing
The LED blinking might be caused by the firmware upgrade that is automatically performed by the driver when you plug the key in the first time.
If, after 20 to 30 seconds, the dongle doesn’t stop blinking, then there is a problem with the dongle's firmware update. To fix the problem:
If that doesn’t help you can:
Analyzing recordings at maximum speed
On some systems and project configurations, an overrun (CPB) or failed overlap (FFT) might occur when analysing data recordings at maximum speed.
To check for these failures you can look in different places:
Migration of PULSE Databases when upgrading from 14.0 or older
On the SQL Installation Troubleshooting page of the Brüel & Kjær support site, you can find information and troubleshooting on the installation and migration of PULSE Databases.
Performance decrease while using a processor with Hyper Threading enabled
Some computers using Intel Hyper Threading (HT) technology might feature lower than expected performance. This may be caused by the overhead induced by the virtualization involved in HT and can be solved by disabling this feature; usually in the computer BIOS.
PULSE Reflex Meta-data
Warning: Meta-data that only differs in upper/lower case text in the name causes an error during save
General PULSE
"Object library invalid or contains references to object definitions that could not be found" message appears while using Microsoft Office
The problem is with VBA when automating Microsoft® Office applications and can be introduced after upgrading to a newer Office version or through a Windows® security update. Affected applications, such as PULSE Balancing Consultant and Sound Power, that use VBA to generate reports will fail to do so. To correct the problem, we suggest that you follow the instructions given here:
If you are using Office Visual Basic for Applications, you may also need to delete the cached versions of the control type libraries. To do this, search your hard drive for ".exd," and delete all the .exd files. The .exd files will be re-created automatically using the new controls the next time that you use VBA.
OLE dispatch error while putting a graph into 64-bit Microsoft® Office 2010
PULSE cannot make reports including Modern Displays (GCX) when using 64-bit Office 2010.
LAN-XI modules do not work with Type 3660-D frame
LAN-XI firmware should be updated to the version that is included with this release to be compatible with Type 3660-D frames.
Anti-virus program issues
Warnings appearing when using PULSE:
A computer security program, like F-Secure®, Norton™ AntiVirus or Symantec™ Endpoint Protection, may report a malicious program attack while PULSE is running. The computer security program is interpreting PULSE's attempts to access system parameters as malicious activity. The computer security program should be configured to ignore these attempts. This is usually achieved by adding the specific PULSE application program to the security program's trusted programs list, or by clicking the "always allow" checkbox in the malicious activity report.
Unexpected program HALT:
Some security programs can interrupt the normal operation of PULSE, causing an unexpected HALT. The security program can, in some cases, even take charge of your computer's CPU process priority, interrupting the running of a PULSE measurement. We recommend to first try the above-mentioned solution or stopping the security program. If nothing else works, you may have to uninstall the security program completely.
Performance degradation:
Anti-virus programs with built-in firewall functionality that scan network packages may cause performance problems affecting network throughput and latency. If the network scan program is required, you may want to detach the PC from the network while running PULSE and use only standard Windows® protection.
Problems starting PULSE Front-end Setup from Time Data Recorder Type 7708
It is possible to start the PULSE Front-end Setup from inside Time Data Recorder. However, you are limited to the type of hardware you can switch between. So when using IDAe frames, you can only change to another IDAe configuration, and when using LAN-XI modules, you can only change to another LAN-XI configuration.
Large numbers of windows cause PULSE to crash or disappear (Windows® XP only)
PULSE may crash or disappear when a large number of windows are open, because the Desktop Heap is too small to contain them all. Expanding the Desktop Heap will solve the problem, but slow PULSE slightly. To expand, follow the steps below:
Windows® 8
PULSE Shortcut Menu
Over the years of PULSE, we have added many different applications, templates and tools to the shortcut menu. Unfortunately, the layout of the Windows® 8 Metro GUI displays all shortcuts simultaneously, making it very difficult to find the correct shortcut. To help fight this problem we have taken the following actions:
For Windows® 8, we have a shortcut to an Explorer with the old Start Menu, making it easy for customers to re-find their shortcuts as in Windows® XP and Windows® 7 if this is preferred.
Windows® 7
Using USB License Dongle with Remote Desktop Connection
When using a USB License Dongle on a PC that is connected to via Remote Desktop Connection, the Dongle will not be recognised by PULSE. This only applies to Remote PCs running Windows® 7. To resolve this problem you should use a license locked to your PC. To rehost your license, please contact your local support representative.
Hint: Using remote control software like Radmin (http://www.radmin.com/) can in many cases be a good solution to this problem.
Saved files disappear
Files saved in some folders, like the "Program Files" folder, may disappear. As a security precaution, Windows® 7 does not allow files to be saved under some folders. Even when the save seems to be completed, the file does not exist. Save the document to another folder.
Vehicle Pass-by Software Type 7788-B and 7788-C
Windows® XP and Windows® 7 have different settings to provide the system from "Front-end connection lost"
Windows® XP
Locate the following entry in the registry:
HKEY_LOCAL_MACHINES\SYSTEM\CurrentControlSet\Services\Tcpip\Parameters
Add a new key (D-word) with the name ‘TcpMaxDataRetransmissions’, and set it to the maximum Hexadecimal value (‘ffffffff’).
Windows® 7
Locate the following entry in the registry:
HKEY_LOCAL_MACHINES\SYSTEM\CurrentControlSet\Services\Tcpip\Parameters
Add a new key (D-word) with the name ‘TcpMaxDataRetransmissions’, and set it to the maximum Hexadecimal value (‘255’).
When using Cisco 1240 series APs powered by POE, make sure no other power supplies are connected! Otherwise this will result in a slow or not responding unit.
The wav file playback button is greyed out in Psychoacoustic Test Bench BZ-5301
While setting up or performing a paired comparison or semantic differential session in Psychoacoustic Test Bench BZ-5301, the buttons for playing the ".wav" files in the playlist may be greyed out.To fix, open the playlist file manually using Jury Test, found in Sound Quality Type 7698's Tools menu, and perform the playback from there.The name and location of the playlist file are found in the Perform Session dialog.
Application | Change Number | Version | Description |
---|---|---|---|
PULSE Reflex Platform, Signal Analysis | PLT-2814 | 18.0.0 | Resolved issue of failure to track slow tacho in Type 8705 |
PULSE Reflex Core | PLT-2755 | 18.0.0 | Resolved issue of failed GADI import |
PULSE Reflex Core | PLT-2724 | 18.0.0 | Meta-data ID lines for I-deas FRF data import are now handled |
PULSE Time Data Recorder | PLT-3010 | 18.0.0 | Resolved issues with global calibration for configurations imported from PULSE LabShop files |
PULSE Reflex Platform, Signal Analysis | PLT-2874 | 18.0.0 | Fixed error when using function statistics on orders |
PULSE Time Data Recorder | PLT-2975 | 18.0.0 | Resolved issue of Type 7708 crashing when exporting to .hdr files |
PULSE Reflex Platform | PLT-2657 | 18.0.0 | Data analysed by I-deas and containing 3-dimensional spectra can now be shown in PULSE Reflex |
PULSE Modal Test Consultant | SD-1133 | 18.0.0 | We now support time files with more than 10 million samples per channel |
PULSE Modal Test Consultant | SD-1140 | 18.0.0 | When generating MIMO FRFs, it is now possible to add or remove DOFs from a sequence after measurements have been completed |
PULSE Modal Test Consultant | SD-1400 | 18.0.0 | Resolved problems reading PTI files that did not contain DOF information |
PULSE Modal Test Consultant | SD-1418 | 18.0.0 | Resolved problem with occasional crash if signals had been removed from the setup prior to export to the OMA application |
PULSE Modal Test Consultant | SD-1419 | 18.0.0 | Added Nutation, Precision and Spin to the DOF Explorer |
PULSE Modal Test Consultant | SD-1423 | 18.0.0 | It is now possible to view shapes with different DOFs and different numbers of DOFs in the Shape Table |
PULSE ODS Test Consultant | SD-1433 | 18.0.0 | Resolved problems with PTI files with large track numbers failing to import into ODSTC. We now accept recordings with track numbers up to 100,000 |
Reflex File Import/Export | STRUICT-1996 | 18.0.0 | We now properly handle UFF files with shapes scaled to Unity Modal Mass |
Reflex File Import/Export | STRUCT-2401 | 18.0.0 | Importing UFF files with double precision results that contain local coordinate systems now works correctly |
Reflex File Import/Export | STRUCT-2543 | 18.0.0 | Exporting geometry containing local coordinate systems now works correctly with double precision UFF files |
Reflex FEM Inport | SD-1420 | 18.0.0 | ODB files support numbering schemes that don't fit within Reflex. We now test to see if that condition is met before renumbering the nodes. This greatly reduces the likelihood of needing to renumber the nodes. |
Reflex Geometry | STRUCT-2455 | 18.0.0 | Implemented ability to deselect a node by reselecting it |
Reflex Geometry | STRUCT-2518 | 18.0.0 | Implemented ability to deselect a coordinate system by reselecting it |
Reflex Modal Analysis | STRUCT-2310 | 18.0.0 | Resolved issue of being unable to perform a global solve after performing a local solve |
Reflex Modal Analysis | STRUCT-2330 | 18.0.0 | Mode Table now updates Complexity values when adding modes and solving for Residues |
Reflex Modal Analysis | STRUCT-2394 | 18.0.0 | Parameter Estimation is now performed in a separate thread so that the application can be more responsive during the solve |
Reflex Modal Analysis | STRUCT-2531 | 18.0.0 | Resolved issue of failure when doing a second Parameter Estimation in the samne setup with a different dataset |
Reflex Correlation | STRUCT-2268 | 18.0.0 | Resolved problems with incorrect Mode Pairings after saving projects |
PULSE Licensing | PD-222 | 18.0.0 | Upgraded FLEXlm to version 11.11.1.2 to support Windows Server 2008 R2 (x64). See Installation troubleshooting above. |
PULSE Installation | PD-223 | 18.0.0 | Fixed issue with failure to install on Windows XP with error 1330 by removing digital signature from Data1.cab. |
PULSE Pass-by | PB-546 | 18.0.0 | Intermediate Values are now stored in the database. |
PULSE Pass-by | PB-541 | 18.0.0 | There is now GADI support for Vaisala weather stations. |
PULSE Pass-by | PB-487 | 18.0.0 | Added constraint for SPL minus Ambient Noise. |
PULSE Pass-by | PB-457 | 18.0.0 | Added constraint for SPL Left minus SPL Right. |
PULSE LabShop and Reflex Order Analysis | PLT-3020 | 17.1.1 | Possible wrong levels or order slices after integrating or differentiating in Modern display. |
Array Acoustics Suite | AAS-1867 | 17.1.1 | Resolved issue of unexpected dialog box appearing when dragging results into the result view. |
PULSE LabShop, CAN Bus Interface | LAB-2925 | 17.1.1 | Problem with using Aux or CAN signals as Tachometer Input has been resolved. |
LAN-XI Software/Firmware | LXISW-1718 | 17.1.1 | Fixed issue with center position of LAN-XI master oscillator that could lead to a deviation of around 2 ppm in absolute frequency precision on any LAN-XI module. |
Array Acoustics Suite | AAS-1736 | 17.1.1 | When showing quasistationary and transient results in mapping, the lower left GCX, sometimes showed wrong data. This has been corrected. No recalculation is needed. |
PULSE Telephone Test | TT-98 | 17.1.1 | Error in VQUEST Macro fixed. Symptom: PESQ and POLQA did not produce any results with low level signals, and the macro "never ended" due to the missing results. |
PULSE LabShop | LAB-2453 | 17.1.1 | Improved stability for with repetitive tests. |
PULSE LabShop | LAB-2901 | 17.1.1 | Fixed issues with playback of multiple tracks. |
PULSE LabShop | LAB-2985 | 17.1.1 | Fixed issue of undefined tacho values in the multi-buffer - values are now put in the multi-buffer as soon as measurement is started. |
PULSE Pass-by | PB-537 | 17.1.1 | It is now possible to load, add and change user-definable procedures. |
PULSE Pass-by | PB-536 | 17.1.1 | The FFT analyzer for the trigger detection will now Accept overload for trigger signal. This to support onboard photocells giving a voltage of 12 volts. |
PULSE Pass-by | PB-533 | 17.1.1 | Fixed issue where sometimes it was not possible to save the run result for stationary measurements. |
PULSE Pass-by | PB-532 | 17.1.1 | It is now possible to delete runs from the result list. This happens when you save the session result to the database. |
PULSE Pass-by | PB-524 | 17.1.1 | Fixed incorrect calculation of MPH. |
PULSE Reflex Telephone Test | TELE-468 | 17.1.1 | Fixed incorrect tolerances in the Reflex Telephone Test PULSE template for test case "Electrical Calibration". |
PULSE Automotive Test Manager | 49479 | 17.1.1 | Fixed unstable update of measurement number when using automatic repetition with Make Rec. File. |
PULSE Structural Dynamics - Modal | 49554 | 17.1.1 | Fixed issue of string overflow error in Geometry Explorer causing crash when doing large numbers of meshes to a single line. |
PULSE Reflex Modal Analysis | 49583 | 17.1.1 | Fixed issue of stack overflow error causing crash when accessing the data shape for ABAQUS ODB files . |
PULSE Reflex Core | 49755 | 17.1.1 | Fixed issue with incorrect dB reference for Power amplitude displays in the data viewer and individual displays for the process chain. |
Contribution Analysis | 49515 | 17.1.1 | Fixed issues with reference point in indoor passby simulation. |
PULSE Noise Source Identification | 47997 | 17.1.1 | Improved handling of transparency in Array Acoustics Suite. |
PULSE Front-end Setup - Analogue Self-test | 49635 | 17.1.1 | Fixed issues with analog self-test crashing if date format was mm/dd/yyyy instead of dd-mm-yyyy. |
PULSE Basic Analyzer | 49498 | 17.1.1 | Fixed issues with multi-buffer indices in playback of multiple tracks. |
HEADQUARTERS: Brüel & Kjær Sound & Vibration Measurement A/S · DK-2850 Nærum · Denmark
Telephone: +45 77 41 20 00 · Fax: +45 45 80 14 05 · www.bksv.com · info@bksv.com
Local representatives and service organisations worldwide